Журнал травм и лечения

Отправить рукопись arrow_forward arrow_forward ..

Medial Migration of the Lag Screw in Gamma Nailing System: A Case Report

Abstract

Issei Nagura, Takako Kanatani, Atsuyuki Inui, Yutaka Mifune, Takeshi Kokubu and Masahiro Kurosaka

Medial lag screw migration in Gamma nail system is an uncommon complication. In 92-year-old woman, the medial lag screw migration into the pelvic with nonunion of trochanteric fracture at 4 months after the osteosynthesis. The computed tomography scan did not reveal intestinal organs injury by the protruded screw and the removal of the implant eased her right hip pain. During the operation, the lag screw migration by the incorrect engagement of the set screw outside of the groove on the lag screw was observed, which was considered the reason. To avoid such a hazardous complication in Gamma nail system for trochanteric fracture, correct engagement of the set screw into the groove on the lag screw should be confirmed after tightening the set screw.